In this study an impedimetric immunosensor was developed in order to determine ciprofloxacin (CIP) in wastewater samples, an emergent contaminant widely found in wastewater. To achieve this, an anti-ciprofloxacin antibody was immobilized on the surface of a printed carbon electrode. Then, the developed immunosensor was applied in wastewater samples from Université Laval residences (Québec, Canada) through the load transfer resistance (<i>R</i> <sub>ct</sub>) using [Fe(CN)<sub>6</sub>]<sup>3-/4-</sup> as a redox probe, and the average CIP concentration was found to be 2.90 × 10<sup>-4</sup> μg mL<sup>-1</sup>. The observed <i>R</i> <sub>ct</sub> changes presented a linear relationship from CIP concentrations of 10<sup>-5</sup> to 1.0 μg mL<sup>-1</sup>, with detection and quantification limits of 2.50 × 10<sup>-6</sup> and 7.90 × 10<sup>-6</sup> μg mL<sup>-1</sup>, respectively. The immunosensor presented high selectivity and repeatability, as well as a good recovery rate in wastewater samples (97%). Significant interference with other compounds was not observed. The proposed method requires only 30 μL of sample without the use of organic solvents or preceding sample preparation and/or extraction techniques. Moreover, the method is fast: only 20 min of incubation followed by 2 min of analysis time was sufficient to obtain the CIP concentration. The method's estimated cost is U$ 2.00 per sample.