Validation of a human immunoturbidimetric assay for measuring canine C-reactive protein

Fabio Gentilini, Danilo Mancini, Francesco Dondi, Laura Ingrà, María Elena Turba, Monica Forni, Paolo Famigli Bergamini

Archivio istituzionale della ricerca (Alma Mater Studiorum Università di Bologna) · 2005 · 13 citations · 0 references

Concepts

Abstract

C-reactive protein (CRP) is an extremely powerful prognostic
\nmarker for a wide variety of diseases. In human medicine
\nimmunoturbidimetric assays (ITAs) are replacing the ELISA test
\ndue to their high degree of accuracy, high-throughput and low
\nturnaround time. Until now, a canine-specific ELISA test has
\nrepresented the gold standard in veterinary medicine for measuring
\nserum CRP. Even though there are contrasting data concerning the
\ncross-reactivity of human and canine CRP, a previous study has
\nnevertheless demonstrated that a commercially available human
\nITA is reliable for measuring canine CRP without method
\nmodification. Since many automated analyzers are furnished with
\nproprietary standardized reagents, the aim of this study was to
\nverify whether a commercially available human ITA (Olympus
\nsystem reagent, CRP ITA, Olympus, Italy) different from the
\nprevious validated assay is also reliable for accurately measuring
\ncanine CRP. The CRP concentrations of 35 canine sera collected from
\neither healthy dogs or dogs with infectious and neoplastic diseases
\nwere assayed by ELISA and ITA on an automated chemistry
\nanalyzer (Olympus AU 400). The ELISA test was carried out
\nfollowing the manufacturer’s instructions while the ITA measurement
\nwas initially performed using the settings and calibration of
\nthe human assay. The results, which were compared using Passing
\nand Bablok regression analysis, showed that the ITA underestimated
\ncanine CRP both with an absolute and a proportional
\ninaccuracy. To overcome this pitfall, the ITA was repeated after
\ncalibration of the automated analyzer with a standard specimen
\nobtained by pooling 3 canine sera with high CRP concentrations.
\nThe standard was assayed 6 times in the same run using the ELISA
\ntest. The CV was 5.8% and the mean concentration of the 6 replicates
\nwas assumed to be the actual concentration of the standard. The
\nmodified assay (ITA-c) was used to evaluate the same 35 serum
\nsamples in duplicate. The results were again compared with the
\nELISA test by Passing and Bablok regression analysis which showed
\na good correlation (y51.18x – 0.69). Moreover, two samples with
\n23.7 mg/dl and 21.7 mg/dl of CRP were serially diluted and
\nassayed in duplicate. The ITA-c was linear in the verified range
\n(y50.9988x þ 0.2 and y50.9982x þ 0.07, respectively). In conclusion,
\nthis study confirms that any commercially available human
\nCRP ITA should be validated before its routine application in dogs.
\nMoreover, the partial cross-reactivity between human and canine
\nCRP allowed us to reliably measure canine CRP after calibration of
\nthe ITA with a standard pooled sample having a known concentration
\nof canine CRP.