Journal of Contemporary Brachytherapy · 2019 · 11 citations · 28 references
MedicineNuclear EngineeringRadiologyRadiation DosimetryRadiation TransportTreatment PlanningAbsorbed DoseBrachytherapy DosimetryMonte Carlo DosimetryPhantom DosimetryRadiation Dose MeasurementRadiation TherapyAnisotropyAnisotropy FunctionBrachytherapy Dose CalculationBrachytherapy Phantom DosimetryGafchromic Film DosimetryMonte Carlo Phantom DosimetryPhantom-based Dose MeasurementRadial Dose Function
PURPOSE: Co (Co0.A86) high-dose-rate (HDR) source in an in-house designed water phantom with egs_brachy Monte Carlo (MC) calculated values. MC results available in the literature were only for unbounded phantoms, and there are no currently published data in the literature for experimental data compared to MC calculations for a bounded phantom. MATERIAL AND METHODS: water phantom using TLD-100 and EBT2 Gafchromic Film and compared it with the egs_brachy results of the same geometry. RESULTS: ,θ), the observed variation with the egs_brachy code calculation is found to be within the acceptable experimental uncertainties of 3%. CONCLUSIONS: ,θ) for the same source using an in-house water phantom. In addition, we validated these experimental results with the values calculated using the egs_brachy MC code, with the same geometry and similar phantom material as used in the experimental methods.