Micromachines · 2019 · 23 citations · 17 references
EngineeringBioengineeringBiomedical EngineeringBiosensorsMicroelectromechanical SystemsSensorsBiomedical SensorsDigestive PhysiologyImplantable MicrodevicesImplantable MicrosystemsMedical InstrumentationPressure SensorsManometryMicrofabricated Capacitive SensorsMiniaturized TransducersPassive SensingPressure SensingResonant FrequencyWireless Pressure SensorsWireless Sensing
We describe a wireless microsystem for gastrointestinal manometry that couples a microfabricated capacitive transducer to a dual-axis inductor, forming a resonant inductor-capacitor (LC) sensor within an ingestible 3D printed biocompatible capsule measuring ø 12 mm × 24 mm. An inductively coupled external telemetry unit wirelessly monitors the pressure dependent resonant frequency of the LC sensor, eliminating the need for integrated power sources within the ingested capsule. In vitro tests in saline show pressure response of -0.6 kHz/mmHg, interrogation distance up to 6 cm, and resolution up to 0.8 mmHg. In vivo functionality is validated with gastrointestinal pressure monitoring in a canine beagle over a 26-hour period.
17
Simple accurate expressions for planar spiral inductances
S.S. Mohan, M. del Mar Hershenson, Stephen Boyd et al. · IEEE Journal of Solid-State Circuits · 1999 · 1.6K citations
Thomas L. Abell, Michael Camilleri, Kevin Donohoe et al. · The American Journal of Gastroenterology · 2007 · 774 citations · Full text
H.A. Wheeler · Proceedings of the IRE · 1942 · 749 citations