Concepedia

Publication | Open Access

Wearable respiration monitoring using an in-line few-mode fiber Mach-Zehnder interferometric sensor

56

Citations

21

References

2019

Year

Abstract

Continuous respiratory monitoring is extensively important in clinical applications. To effectively assess respiration rate (RR), tidal volume (TV), and minute ventilation (MV), we propose and experimentally demonstrate a respiration monitoring system using an in-line few-mode fiber Mach-Zehnder interferometer (FMF-MZI), which is the first to introduce in-line MZI into an optimal wearable design for respiration rate and volume monitoring. The optimal linear region of the proposed sensor is analyzed and positioned by a flexible arch structure with curvature sensitivity up to 8.53 dB/m<sup>-1</sup>. Respiration monitoring results are in good agreement with a standard spirometer among different individuals. The difference in TV estimation is ± 0.2 L, and the overall error of MV estimation is less than 5%.

References

YearCitations

Page 1