Physics and Imaging in Radiation Oncology · 2019 · 16 citations · 32 references
MedicineOncologyRadiation OncologyClinical Radiation OncologySurgical OncologyTreatment PlanningDeep Inspiration Breath HoldExternal Beam RadiotherapyMotion ManagementRadiation TherapyRadiation Treatment PlanningStereotactic RadiotherapyBreath-hold RadiotherapyPatient AlignmentReal-time Motion TrackingReal-time Patient TrackingReal-time TrackingSetup Time MeasurementTreatment ParametersTumor Motion Management
BACKGROUND AND PURPOSES: Motion management is crucial for optimal stereotactic body radiotherapy (SBRT) of moving targets. We aimed to describe our clinical experience with real-time tracking of lung-specific electromagnetic transponders (EMTs) for SBRT of early stage non-small cell lung cancer in free-breathing (FB) or deep inspiration breath-hold (DIBH). MATERIAL AND METHODS: Seven patients were implanted with EMTs. Simulation for SBRT was performed in FB and in DIBH. We prescribed 60 Gy in 3, 5 or 8 fractions to the tumor and delivered SBRT with volumetric modulated arcs and a 6 MV flattening filter free photon beam. Patients' setup at the linac was performed using EMT positions and cone-beam CT (CBCT) verification. Four patients were treated in DIBH because of a dosimetric benefit. We analysed patient alignment and treatment delivery parameters using DIBH or FB and EMT real-time tracking. RESULTS: There were no complications from the EMT implantation. Visual inspection of CBCT before and/or after SBRT revealed good alignment of structures and EMTs. The median setup time was 9.8 min (range: 4.6-34.1 min) and the median session time was 14.7 min (range: 7.3-36.5 min). EMT positions in lungs remained stable during overall treatment and allowed real-time tracking both in FB and in DIBH SBRT. The treatment beam was gated when EMT centroid position exceeded tolerance thresholds ensuring correct delivery of radiation to the tumor. CONCLUSION: Using EMTs for real-time tracking of tumor motion during lung SBRT proved to be safe, accurate and easy to integrate clinically for treatments in FB or DIBH.
32
Hiroki Shirato, Kei Kitamura, Shinichi Shimizu et al. · International Journal of Radiation Oncology*Biology*Physics · 2002 · 1.4K citations
Patrick A. Kupelian, Twyla R. Willoughby, Arul Mahadevan et al. · International Journal of Radiation Oncology*Biology*Physics · 2006 · 488 citations
Twyla R. Willoughby, Patrick A. Kupelian, Jean Pouliot et al. · International Journal of Radiation Oncology*Biology*Physics · 2006 · 426 citations