IEEE Transactions on Instrumentation and Measurement · 2000 · 49 citations · 6 references
EngineeringElectrical EngineeringAnalog ElectronicsInstrumentationPower ElectronicsAc-dc ConversionAc-dc ConvertersCalibrationDc-dc ConvertersElectrical MeasurementEquipment CalibrationDc-ac ConversionPower MeasurementPower Quality In Ac-dc ConversionPrecision MeasurementVoltage Measurement
Three methods for the determination of the RMS value of AC voltage at low frequencies are compared: the step calibration of a digitally synthesized source, the optimized sampling method, and the AC-DC voltage transfer with a multijunction thermal converter. The three methods agree within 1/spl times/10/sup -6/ with an uncertainty of 2.5/spl times/10/sup -8/ in the frequency range from 10 to 100 Hz.
6