IEEE Transactions on Instrumentation and Measurement · 2006 · 43 citations · 8 references
EngineeringElectrical EngineeringAnalog ElectronicsInstrumentationPower ElectronicsAc-dc ConvertersAnalog InstrumentationCalibrationCurrent-voltage CharacteristicsElectrical MeasurementEquipment CalibrationCalibration Techniques For Analog CircuitsCurrent Measurement Data ProcessingCurrent Sensor CalibrationUniversal Serial BusVoltage Reference Circuits
This paper describes a new ac-dc current-shunt calibration system with a current range of 1 mA-100 A and a frequency range from 10 Hz to 100 kHz. The new system consists of an ac-dc comparator controlled via a universal serial bus and a set of reference coaxial shunts with dedicated thermal voltage converters. The design of the custom-built shunts has been optimized through a theoretical modeling. The results of the calibrations of the reference coaxial shunts are presented.
8