Journal of Bone and Joint Surgery - British Volume · 1997 · 494 citations · 20 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryJoint Replacement SurgeryProstheticsShoulder SurgeryLower-limb ProstheticsProsthetic AdjustmentProsthetic Component SelectionProsthetic FabricationProsthetic FittingUpper-limb ProstheticsAdaptabilityAnatomical RestorationArticular Surface AnatomyCustomization Of Prosthetic ComponentsHumeral AnatomyHumeral Component DesignHumeral Head AnatomyInclination Variations
The study aims to restore the proximal humerus’s 3‑D geometry by developing a modular, anatomically adaptable humeral component. The authors mapped proximal humerus geometry on cadavers using a digitised measuring device and designed a modular prosthesis that adapts to individual inclination, retroversion, medial offset, and posterior offset. The study found that proximal humeral shape and orientation vary widely, and existing prostheses’ fixed geometry fails to accommodate these variations, limiting anatomical restoration.
We have studied the three-dimensional geometry of the proximal humerus on human cadaver specimens using a digitised measuring device linked to a computer. Our findings demonstrated the variable shape of the proximal humerus as well as its variable dimensions. The articular surface, which is part of a sphere varies individually in its orientation as regards inclination and retroversion, and it has variable medial and posterior offsets. These variations cannot be accommodated by the designs of most contemporary humeral components. Although good clinical results can be achieved with current modular and non-modular components their relatively fixed geometry prevents truly anatomical restoration in many cases. To try to restore the original three-dimensional geometry of the proximal humerus, we have developed a new type of humeral component which is modular and adaptable to the individual anatomy. Such adaptability allows correct positioning of the prosthetic head in relation to an individual anatomical neck, after removal of the marginal osteophytes. The design of this third-generation prosthesis respects the four geometrical variations which have been demonstrated in the present study. These are inclination, retroversion, medial offset and posterior offset.
20
Articular Geometry of the Glenohumeral Joint
Louis J. Soslowsky, Evan L. Flatow, Louis U. Bigliani et al. · Clinical Orthopaedics and Related Research · 1992 · 342 citations