The Determination of Cardiac Output by a Continuous Recording System Utilizing Iodinated (I <sup>131</sup> ) Human Serum Albumin
43
Citations
5
References
1951
Year
Heart FailureBiomedical EngineeringDiastolic FunctionBlood FlowCardiac OutputDilution CurveRotameter MethodClinical ChemistryLaboratory MedicineCardiologyBlood Flow MeasurementCardiac MechanicRadiologyHealth SciencesCardiovascular ImagingCardiovascular DiseasePhysiologyHuman Serum AlbuminElectrophysiologyCardiovascular PhysiologyMedicineAnesthesiology
A method is outlined for the determination of cardiac output by the injection of iodinated (I 131 ) human serum albumin and continuous recording of the dilution curve. The calibration of the flow system by means of the final dilution and blood volume is discussed along with the possible application to an external measuring method. The results of eight dilution curves obtained from dogs are presented and the calculated cardiac outputs compared with the values measured by the rotameter method.