Clinical outcomes of pallidal deep brain stimulation for dystonia implanted using intraoperative MRI
13
Citations
28
References
2019
Year
Intraoperative MRI-guided lead placement in patients with dystonia showed improvement in clinical outcomes comparable to previously reported results using awake MER-guided lead placement. The accuracy of lead placement was high, and the procedure was well tolerated in the majority of patients. However, a number of patients experienced serious adverse events that were attributable to the introduction of a novel technique into a busy neurosurgical practice, and which led to the revision of protocols, product inserts, and on-site training.