Publication | Closed Access
Predicting Thrombosis Formation in 1-mm-Diameter Arterial Anastomoses with Transit-Time Ultrasound Technology
25
Citations
17
References
2017
Year
At 20 minutes postoperatively, blood flows greater than 0.30 ml/minute are highly suggestive of patency, and flows less than 0.21 ml/minute are highly suggestive of failure. The authors therefore recommend a minimal cutoff flow value of 0.30 ml/minute for vessels ranging from 0.6 to 1.2 mm in diameter to predict long-term postoperative vascular patency.