A Solid-State Microdosimeter for Dose and Radiation Quality Monitoring for Astronauts in Space

Stefania Peracchi, Naruhiro Matsufuji, Angela Kok, Marco Povoli, Michael Jackson, Anatoly Rosenfeld, Linh T. Tran, Benjamin James, David Bolst, Dale A. Prokopovich,

IEEE Transactions on Nuclear Science · 2019 · 15 citations · 13 references

DOIFull text

Open access

Concepts

Abstract

This article presents a study of the response of the silicon on insulator (SOI) microdosimeter with 3-D sensitive volumes (SVs) to 400 MeV/u <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">16</sup> O and 500 MeV/u <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">56</sup> Fe ions mimicking galactic cosmic rays outside and inside the International Space Station (ISS). An average quality factor ( <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\bar {Q}$ </tex-math></inline-formula> ) and the dose equivalent ( <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$H$ </tex-math></inline-formula> ) of the radiation field were obtained experimentally, and the results were compared with GEANT4 simulations.

References

13