Interference of lipemia in samples for routine coagulation testing using a Sysmex CS‐5100 coagulometer
13
Citations
6
References
2019
Year
For PT and antithrombin activity a mean percentual difference between the two determinations lower than within-subject biologic variation and one of the Fraser's quality specifications can be considered as a confounding 'noise' factor that is neither analytically nor clinically relevant. If the instrument determines a result on the first run, for PT, aPTT, D-dimer concentration and antithrombin activity tests, the double plasma high-speed centrifugation is unnecessary. It is instead necessary if fibrinogen >2 g/L or if the instrument cannot determine a result on the first run.