Limitations and Precautions in the Use of the Farnsworth‐Munsell Dichotomous D‐15 Test

28

Citations

30

References

2019

Year

Abstract

All these factors affect the outcome of the test and need specification and implementation if the test is to be applied consistently and equitably. We recommend the following: practitioners should never rely on a single color vision test regardless of the color vision standard; lighting should be Tcp '' 6500 K and Ra > 90; illuminance levels should be between 200 and 300 lux if detection of color vision deficiency is a priority or between 300 and 1000 lux if the need is to test at the level where illuminance has minimal influence on performance; illuminance should be reported; time limits should be set between 1 and 2 minutes; repeat testing (beyond the specified test and one retest) should be carried out only with authorization; and initial and repeated results should be reported. A set of test instructions to assist in the consistent application of the test is provided in the Appendix.