Concepedia

Publication | Open Access

Repeatability and reproducibility of measurements of low dissolved radiocesium concentrations in freshwater using different pre-concentration methods

10

Citations

19

References

2019

Year

Abstract

We assessed the repeatability and reproducibility of methods for determining low dissolved radiocesium concentrations in freshwater in Fukushima. Twenty-one laboratories pre-concentrated three of 10 L samples by five different pre-concentration methods (Prussian-blue-impregnated filter cartridges, coprecipitation with ammonium phosphomolybdate, evaporation, solid-phase extraction disks, and ion-exchange resin columns), and activity of radiocesium was measured. The z-scores for all of the 137Cs results were within ± 2, indicating that the methods were a good degree of precision. The relative standard deviations (RSDs) indicating the reproducibility among different laboratories were larger than the RSDs indicating the repeatability in each separate laboratory.

References

YearCitations