The Analyst · 2019 · 63 citations · 34 references
PharmacyPharmaceutical AnalysisPharmacologyAnalytical ChemistryAnalytical ToxicologyBiosensorsBioanalytical Method ValidationClinical AnalysisInstrumental AnalysisSpectroscopic Analysis Of PharmaceuticalsBlood Drug AnalysisPoint-of-care Drug AssaysPoint-of-care Drug TestingSaliva BiosensingSalivary Analyte DetectionSalivary Metabolite Detection
A simple-to-use device to measure drugs in saliva, blood plasma, and whole blood for point-of-care analysis and treatment of overdose patients has been investigated. A rudimentary flow strip has been developed to separate opioids from these biofluids for analysis by surface-enhanced Raman spectroscopy (SERS). The strips are based on lateral flow assays, in which the antibodies have been substituted by SERS-active pads for detection. Samples of codeine and fentanyl, artificially added to these biofluids, were measured using the strips by a field-usable Raman spectrometer. We report measurement of these drugs in these biofluids from 0.5 to 5 μg mL-1 in 5 minutes. Calculated limits of detection for the spectra suggest that these drugs could be measured at 5 to 20 ng mL-1 with improvements in the strips' separation capability.
34
Surface raman spectroelectrochemistry
David L. Jeanmaire, Richard P. Van Duyne · Journal of Electroanalytical Chemistry · 1977 · 4.7K citations
The composition of body tissues
Helen Q. Woodard, D. R. White · British Journal of Radiology · 1986 · 761 citations
Agriculture, Anatomy, Physiology +13
Characteristic Raman frequencies of organic compounds
W.J.O.-T · Journal of Molecular Structure · 1974 · 736 citations
Natural Sciences, Chemistry, Physics +14
Alana M. Vivolo‐Kantor, Puja Seth, R. Matthew Gladden et al. · MMWR Morbidity and Mortality Weekly Report · 2018 · 433 citations · Full text