Physics in Medicine and Biology · 2019 · 40 citations · 31 references
MedicineMaterials ScienceRadiation DosimetryRadiation PhysicsRadiobiologyAbsorbed DoseHigh-dose Radiation EffectsMonte Carlo DosimetryMonte Carlo Radiation TransportRadiation Dose MeasurementRadiobiological ModelingEnergy DepositionLocal EnhancementMonte Carlo Dose ModelingMonte Carlo Dose SimulationNanoparticlesRadiobiological EffectivenessRadiobiological EffectsRadiobiological Response
Nanoparticles (NPs) containing high atomic number (high-Z) materials have been shown to enhance the radiobiological effectiveness of ionizing radiation. This effect is often attributed to an enhancement of the absorbed dose in the vicinity of the NPs, based on Monte Carlo simulations that show a significant local enhancement of the energy deposition on the microscopic scale. The results of such simulations may be significantly biased and lead to a severe overestimation of the dose enhancement if the condition of secondary particle equilibrium is not met in the simulation setup. This current work shows an approach to estimate a 'realistic' dose enhancement from the results of such biased simulations which is based on published photon interaction data and provides a way for correcting biased results.
31
BEAM: A Monte Carlo code to simulate radiotherapy treatment units
D. W. O. Rogers, Bruce Faddegon, G Ding et al. · Medical Physics · 1995 · 1.4K citations