Publication | Closed Access
Magnetization and screening current in an 800 MHz (18.8 T) REBCO nuclear magnetic resonance insert magnet: experimental results and numerical analysis
89
Citations
50
References
2019
Year
Numerical AnalysisSuperconducting MaterialEngineeringMagnetic ResonanceRebco InsertMagnetic MaterialsMagnetoresistanceMagnetismSuperconductivityElectrical EngineeringHigh-tc SuperconductivityPhysicsMagnetic MeasurementRebco MagnetsExperimental ResultsNatural SciencesScreening Current EffectMagnetic DeviceMagnetic FieldMagnetic Property
The nonuniform superconducting current distribution in REBCO coated conductor, including a varying-field-induced screening current, is responsible for a significant magnetization effect that not only degrades the field quality of REBCO magnets, but also introduces risks of overstressing conductor. This paper presents our experimental and simulation studies on the screening current effect on an 800-MHz (18.8-T) REBCO insert (H800) that together with a 500-MHz LTS NMR magnet (L500) constitutes the MIT 1.3-GHz LTS/HTS NMR magnet (1.3G). To develop our simulation model, which subsequently validated by a good agreement between simulation and experiment, we chose H800, Coil 1 of the 3-coil assembly operated alone and the entire H800, for the sources of experimental data, specifically their remnant fields after current discharge and diminished axial fields during operation. Armed with this valid model, we examined in detail the negative effects of screening current on H800, an important 1.3G component. Our simulation indicates that the screening current, nonuniformly distributed in the REBCO conductor, not only deteriorates H800 field, both strength and homogeneity, thus that of 1.3G, but also may overstress the REBCO conductor.
| Year | Citations | |
|---|---|---|
Page 1
Page 1