An investigation of quantitative accuracy for deep learning based denoising in oncological PET

Wenzhuo Lu, John A. Onofrey, Yihuan Lu, Luyao Shi, Tianyu Ma, Yaqiang Liu, Chi Liu

Physics in Medicine and Biology · 2019 · 131 citations · 41 references

Concepts

Abstract

bias. Fully 3D U-net is superior to several existing denoising methods, including Gaussian filter, anatomical-guided non-local mean (NLM) filter, and MAP reconstruction with Quadratic prior and relative difference prior, in terms of superior image quality and trade-off between noise and bias. Furthermore, incorporating aligned CT images has the potential to further improve the quantitative accuracy in multi-channel U-net. We found the optimal architectures and parameters of deep learning based methods are different for absolute quantitative accuracy and visual image quality. Our quantitative results demonstrated that fully 3D U-net can both effectively reduce image noise and control bias even for sub-centimeter small lung nodules when generating standard dose PET using 10% low count down-sampled data.