Current phototherapy practice on Java, Indonesia

Mahendra Tri Arif Sampurna, Kinanti Ayu Ratnasari, Darto Saharso, Arend F. Bos, Pieter J.J. Sauer, Peter H. Dijk, Christian V. Hulzebos

BMC Pediatrics · 2019 · 12 citations · 22 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND: In Indonesia, the burden of severe hyperbilirubinemia is higher compared to other countries. Whether this is related to ineffective phototherapy (PT) is unknown. The aim of this study is to investigate the performance of phototherapy devices in hospitals on Java, Indonesia. METHODS: In 17 hospitals we measured 77 combinations of 20 different phototherapy devices, with and without curtains drawn around the incubator/crib. With a model to mimic the silhouette of an infant, we measured the irradiance levels with an Ohmeda BiliBlanket Meter II, recorded the distance between device and model, and compared these to manufacturers' specifications. RESULTS: = 0.1838; P < 0.05). The effect of curtains on irradiance levels was highly variable, ranging from - 6.15 to + 15.4 μW/cm2/nm, with a mean difference (SD) of 1.82 (3.81) μW/cm2/nm (P = 0.486). CONCLUSIONS: In half of the hospitals that we studied on Java the levels of irradiance are too low and, in some cases, too high. Given the risks of insufficient phototherapy or adverse effects, we recommend that manufacturers provide radiometers so hospitals can optimize the performance of their phototherapy devices.

References

22