Concepedia

Publication | Open Access

Absolute calibration of GafChromic film for very high flux laser driven ion beams

18

Citations

33

References

2019

Year

Abstract

We report on the calibration of GafChromic HD-v2 radiochromic film in the extremely high dose regime up to 100 kGy together with very high dose rates up to 7 × 10<sup>11</sup> Gy/s. The absolute calibration was done with nanosecond ion bunches at the Neutralized Drift Compression Experiment II particle accelerator at Lawrence Berkeley National Laboratory (LBNL) and covers a broad dose dynamic range over three orders of magnitude. We then applied the resulting calibration curve to calibrate a laser driven ion experiment performed on the BELLA petawatt laser facility at LBNL. Here, we reconstructed the spatial and energy resolved distributions of the laser-accelerated proton beams. The resulting proton distribution is in fair agreement with the spectrum that was measured with a Thomson spectrometer in combination with a microchannel plate detector.