Alanine dosimetry at NPL - the development of a mailed reference dosimetry service at radiotherapy dose levels

29

Citations

3

References

1998

Year

Abstract

In this paper we describe the work that has been carried out at NPL to develop a mailed alanine reference dosimetry service for radiotherapy dose levels. The service is based on alanine / paraffin wax dosimeters produced at NPL. Using a data analysis technique based on spectrum fitting, it has been possible to achieve a precision of dose measurement better than ±0.05 Gy (1 o). A phantom set has been developed for use in high energy photon beams, which enables simultaneous irradiation of alanine dosimeters and ionisation chambers in a well defined geometry. Studies in photon beams of energies between Co-60 and 20 MeV have shown no significant energy dependence (<1%) for alanine relative to dose determination using a graphite calorimeter. Work is underway to extend the service to electron beams, and preliminary results are presented on the direct calibration of alanine in electron beams using a graphite calorimeter.