Comparison of Three Different Types of Two-Implant-Supported Magnetic Attachments on the Stress Distribution in Edentulous Mandible

Fengling Hu, Yiming Gong, Zhen Bian, Xiaoying Zhang, Bin Xu, Jianguo Zhang, Xiaojun Shi, Youcheng Yu, Liang Song

Computational and Mathematical Methods in Medicine · 2019 · 14 citations · 32 references

DOIFull text

Open access

Concepts

Abstract

Two-implant-retained mandibular overdentures with magnetic attachments can provide an effective treatment modality for edentulous patients. In this study, a three-dimensional finite element analysis was used to compare the biomechanical characteristics of three different types of magnetic attachments in two-implant-retained mandibular overdentures. Flat-type, dome-type, and cushion-type of the magnetic attachments were designed to retain the overdenture. Four types of load were applied to the overdenture in each model: 100 N vertical and oblique loads on the right first molar and a 100 N vertical load on the right canine and the lower incisors. The biomechanical behaviors of peri-implant bone, abutment, and mucosa were recorded. In vertical incisors, vertical right canine, and oblique molar loading condition, the flat-type group exhibited the highest levels of maximum equivalent strain/stress in the peri-implant bone. The total deformation of mucosa and the maximum equivalent strain/stress in the oblique molar loading condition are about two times as the vertical molar loading condition. These results suggested that both cushion-type and dome-type of the magnetic attachments are better choices in two-implant-retained mandibular overdentures, and oblique loading is more harmful than vertical loading.

References