High-resolution high-speed microwave-multiplexed low temperature microcalorimeters for the HOLMES experiment

Bradley K. Alpert, Daniel Becker, D. Bennet, M. Biasotti, M. Borghesi, G. Gallucci, M. De Gerone, M. Faverzani, E. Ferri, Joseph W. Fowler,

The European Physical Journal C · 2019 · 23 citations · 27 references

DOIFull text

Open access

Concepts

Abstract

We present the first performance results obtained with microwave multiplexed Transition Edge Sensors prototypes specifically designed for the HOLMES experiment, a project aimed at directly measuring the electron neutrino mass through the calorimetric measurement of the $$^{163}$$ Ho electron capture spectrum. The detectors required for such an experiment feature a high energy resolution at the Q–value of the transition, around $$\sim $$ 2.8 keV, and a fast response time combined with the compatibility to be multiplexed in large arrays in order to collect a large statistics while keeping the pile-up contribution as small as possible. In addition, the design has to be suitable for future ion-implantation of $$^{163}$$ Ho. The results obtained in these tests allowed us to identify the optimal detector design among several prototypes. The chosen detector achieved an energy resolution of (4.5 ± 0.3) eV on the chlorine K $$_\alpha $$ line, at $$\sim $$ 2.6 keV, obtained with an exponential rise time of 14 $$\upmu $$ s. The achievements described in this paper pose a milestone for the HOLMES detectors, setting a baseline for the subsequent developments, aiming to the actual ion-implantation of the $$^{163}$$ Ho nuclei. In the first section the HOLMES experiment is outlined along with its physics goal, while in the second section the HOLMES detectors are described; the experimental set-up and the calibration source used for the measurements described in this paper are reported in Sects. 3 and 4, respectively; finally, the details of the data analysis and the results obtained are reported in Sect. 6.

References

27