Journal of Instrumentation · 2019 · 58 citations · 3 references
EngineeringElectrical EngineeringNuclear EngineeringPhotonicsRadiation DetectionSemiconductor DevicesDetector TechnologyEnergy ResolutionIntegrated PhotonicsRadiation DetectorsSemiconductor DetectorsSilicon PhotonicsCharge IntegrationDetector Energy ResolutionDigitizationExperimental CharacterizationRadiation Energy ResolutionSemiconductor Radiation DetectorsSilicon PhotomultipliersTime Resolution
We present the experimental characterization of the TOFPET2, a readout and digitization ASIC for radiation detectors using Silicon Photomultipliers. The circuit is designed in CMOS 110 nm technology, has 64 independent channels and is optimized for time-of-flight measurement in PET or other applications. The chip has quad-buffered TDCs and charge integration QDCs in each channel. The Coincidence Time Resolution (CTR) of 511 keV photon pairs from a 22Na point source measured with 2 × 2 × 3 mm3 LSO:Ce crystals co-doped with 0.2% Ca is 118 and 119 ps FWHM when using respectively the SiPMs NUVHD 40um from Fondazione Bruno Kessler (FBK) and the S14160-3050HS MPPC from Hamamatsu Photonics (HPK). The energy resolution obtained for the 511keV photopeak is 10.5 and 12% FWHM when using respectively the SiPMs PM3325-WB from KETEK and the QFBR-S4N44P164S from Broadcom Inc.
3
TOFPET ASIC for PET applications
M. Rolo, R. Bugalho, F.M. Gonçalves et al. · Journal of Instrumentation · 2013 · 124 citations · Full text