Publication | Closed Access
Impacts of respiratory phase shifts on motion‐tracking accuracy of the CyberKnife Synchrony™ Respiratory Tracking System
27
Citations
27
References
2019
Year
We have investigated the motion-tracking accuracy of the CyberKnife SRTS for cases with the respiratory phase shift between the target and the LED marker. The maximum tracking errors for 90% probability were within 2.4 mm when the target delays to the LED markers. When LED marker delays, however, very large tracking errors were observed. With a long respiratory cycle, the tracking errors were greatly improved to less than 2.2 mm. Coaching slow breathing will be useful for accurate motion tracking radiotherapy.