Concepedia

Publication | Closed Access

Clinical Evaluation of a New Swept-Source Optical Coherence Biometer That Uses Individual Refractive Indices to Measure Axial Length in Cataract Patients

59

Citations

22

References

2019

Year

Abstract

In eyes with long axial length, when the conventional IOL calculation was optimized with the equivalent refractive index-based instrument, we need to consider that IOL calculation using OCTB1 tends to cause slightly myopic refractive prediction error.