Intraoral Digital Impression Technique for a Neonate With Bilateral Cleft Lip and Palate

Jilen Patel, John C. Winters, Mark Walters

The Cleft Palate-Craniofacial Journal · 2019 · 44 citations · 9 references

Concepts

Abstract

OBJECTIVE: Dental casts are an important aspect in the treatment planning, documentation, and analysis of the dental arch forms of infants with cleft lip and palate (CLP). Impression taking in a neonate is a technique-sensitive procedure, which can result in foreign body dislodgement and carries a small risk of aspiration and airway obstruction. The advent of digital dental technologies and intraoral scanning may facilitate safer, more effective, and accurate impressions for CLP infants. DESIGN: A digital intraoral scan of a 3-month-old with bilateral CLP (BCLP) was compared with a conventional alginate impression taken prior to primary lip repair. SETTING: Princess Margaret Hospital for Children. MAIN OUTCOME MEASURES: To test the applicability and accuracy of digital impression taking for a neonate with BCLP and palate using digital morphometrics. RESULTS: The average deviation of points ranged above and below the plane of superimposition from +0.78 mm to -0.42 mm with a maximum range of +2.80 mm to -2.80 mm and standard deviation of 0.88 mm. The premaxillary segment showed the greatest degree of variation. CONCLUSIONS: This is to our knowledge the first report that illustrates the use of a digital impression system to scan the oral structures of a neonate with BCLP. Digital scanning was found to be fast, accurate, and safe, when compared to a conventional alginate impression technique.

References

9