11.4 A Fast-Readout Mismatch-Insensitive Magnetoresistive Biosensor Front-End Achieving Sub-ppm Sensitivity

Xiahan Zhou, Michael Sveiven, Drew A. Hall

2019 · 10 citations · 5 references

Concepts

Abstract

Magnetic biosensors have drawn significant research interest as a means to replace large optical instrumentation commonly found in centralized diagnostic laboratories. Magnetic biosensing is attractive since biological samples lack a magnetic background thus enabling high sensitivity measurements. Figure 11.4.1 shows a magnetic immunoassay where: 1) the surface of a magnetic sensor is coated with capture molecules (e.g., antibodies) specific to the biomarker (e.g., antigen) of interest, 2) the sample, detection molecules, and magnetic nanoparticles (MNPs) are added and self-assemble thus tethering the MNPs to the sensor, and 3) the sensor is read out by applying a magnetic field, H <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">A</sub> , that magnetizes the superparamagnetic MNPs producing a stray field that perturbs the local magnetic field. The underlying magnetoresistive (MR) sensor detects this change in field as a miniscule ppm change in resistance, R <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">sig</sub> , that is directly proportional to the number of MNPs. Unfortunately, Rsig is superimposed on the baseline resistance, R <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">0</sub> , resulting in a large baseline-to-signal ratio (R <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">0</sub> /R <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">sig</sub> >10,000) that requires a very high dynamic range (DR>120dB) front-end for quantitative detection [1,2].

References

5