Journal of Medical Physics · 2019 · 10 citations · 35 references
MedicineBiomedical EngineeringMaterials ScienceNanomaterialsNanotechnologyRadiation DosimetryDetector TechnologyExternal Beam DosimetryOptically Stimulated Luminescence DosimetryPhoton DosimetrySmall Field DosimetryStereotactic RadiosurgeryClinical Photon Beam DosimetryDetectorsMedical Dosimetry ApplicationsMetal-oxide-semiconductor Field-effect Transistor DosimetryOptically Stimulated Luminescence DosimetersOutput Factor MeasurementPhoton Small Field DosimetryStereotactic Radiosurgery Dosimetry
<b>Aim:</b> The aim of this study was to investigate the adequacy of nanoDot optically stimulated luminescence (OSL) dosimeter for small field dosimetry before its <i>in vivo</i> applications in CyberKnife SRS unit. <b>Materials and Methods:</b> A PTW 60018 SRS Diode, 60019 microDiamond, and Gafchromic EBT3 films were used along with a nanoDot carbon-doped aluminum oxide OSL dosimeter to collect and compare beam data. In addition, the EGSnrc/BEAMnrc code was employed to simulate 6-MV photon beams of CyberKnife SRS system. <b>Results:</b> All detectors showed good consistency with each other in output factor measurements for cone sizes of 15 mm or more. The differences were maintained within 3% for these cones. However, OSL output factors showed higher discrepancies compared to those of other detectors for smaller cones wherein the difference reached nearly 40% for cone size of 5 mm. Depending on the performance of OSL dosimeter in terms of output factors, percentage depth doses (PDDs) were only measured for cones equal to or larger than 15 mm. The differences in PDD measurements were within 5% for depths in the range of 5–200 mm. <b>Conclusion:</b> Its low reliable readings for cones smaller than 15 mm should be considered before its <i>in vivo</i> applications of Cyberknife system.
35
BEAM: A Monte Carlo code to simulate radiotherapy treatment units
D. W. O. Rogers, Bruce Faddegon, G Ding et al. · Medical Physics · 1995 · 1.4K citations
Small fields: Nonequilibrium radiation dosimetry
Indra J. Das, G Ding, Anders Ahnesjö · Medical Physics · 2007 · 663 citations
A new formalism for reference dosimetry of small and nonstandard fields
Rodolfo Alfonso, Pedro Andreo, R. Capote et al. · Medical Physics · 2008 · 550 citations