Publication | Open Access
A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control
35
Citations
21
References
2019
Year
Machine performance check was verified as a daily quality assurance tool to check machine output and symmetry while assessing against an independent detector on a weekly basis. MPC output detection can be improved by regular SPC-based trend analysis to measure drifts in the inherent device and control systematic and random variations thereby increasing confidence in its capabilities as a QA device. A 3-monthly MPC calibration assessment was recommended based on SPC capability and acceptability calculations.
| Year | Citations | |
|---|---|---|
Page 1