The colonization rate of needleless connector and the impact of disinfection for 15 s on colonization: A prospective pre- and post-intervention study

İlker Devrim, Nevbahar Demiray, Yeliz Oruç, Kenan Sipahi, İlknur Çağlar, Ferhat Sarı, Nuriye Turgut, Gülhan Atakul, Nihal Özdamar, Vecihe Dursun,

The Journal of Vascular Access · 2019 · 14 citations · 15 references

Concepts

Abstract

PURPOSE: The optimal scrubbing time for the disinfection of the surface of needleless connectors has not been determined. The evidence for successful needleless connector decontamination with 70% isopropyl alcohol ranges from 5 to 60 s. The aim of this prospective study was to identify colonization on the external surface of needleless connectors on central venous catheters and measure the efficiency of 15 s of scrubbing with 70% alcohol. METHOD: A total of 31 patients were included. Samples were collected adhering to aseptic no-touch technique policies. Two samples were collected from the injectable surface of the needleless connector with sterile sodium chloride 0.9% moistened and a dry swab from the same site. Then the surface was scrubbed with alcohol 70% for 15 s for disinfection and second couples of samples for the cultures were taken after disinfection. FINDINGS: was responsible for all the colonization. After disinfection for 15 s, no isolation exceeding 1 CFU/plate was present. CONCLUSION: Scrubbing for 15 s with 70% alcohol was found to be successful at elimination of colonization of the surface of needleless connector even in high microorganism counts.