Validation of post‐treatment PET‐based dosimetry software for hepatic radioembolization of Yttrium‐90 microspheres

Nichole M. Maughan, José Garcia-Ramirez, Matt Arpidone, Amy Swallen, Richard Laforest, S Goddu, Parag J. Parikh, Jacqueline E. Zoberi

Medical Physics · 2019 · 33 citations · 21 references

Abstract

Dose calculation results from MIM SurePlan were found to be in good agreement with our in-house algorithm. This indicates that MIM SurePlan performs as it should and, hence, can be deemed accepted and commissioned for clinical use for post-implant PET-based dosimetry of <sup>90</sup> Y radioembolization.