Comparison of Tin Filter–Based Spectral Shaping CT and Low-Dose Protocol for Detection of Urinary Calculi

Amirhossein Mozaffary, Tugce Agirlar Trabzonlu, Donald Kim, Vahid Yaghmai

American Journal of Roentgenology · 2019 · 39 citations · 17 references

Concepts

Abstract

OBJECTIVE: The purpose of this study was to assess the performance of tin filter-based spectral shaping CT compared with routine low-dose CT for detection of urolithiasis. MATERIALS AND METHODS: ) and size-specific dose estimate (SSDE) were recorded. To prospectively compare the performances of the spectral shaping protocol (Sn150 kV) with the standard (120 kV) and routine low-dose (100 kV) protocols, a phantom (sheep kidneys) containing stones were also scanned with each protocol and evaluated by two radiologists. RESULTS: compared with CT performed with routine low-dose and standard-dose protocols (p < 0.05). Accordingly, it also led to 24% and 55% reduction in SSDE compared with the low-dose and standard protocols (p < 0.05). Subjective image quality and signal-to-noise ratio were similar between the tin filtration and the routine low-dose groups (p > 0.05). The objective image noise was similar in the three groups (p > 0.05). The phantom study showed no difference in detection of renal stones between the three tube potential settings. CONCLUSION: Using spectral shaping with tin filtration can substantially reduce radiation dose compared with routine standard- and low-dose abdominal CT for urinary stone disease.