Thrombogenicity assessment of Pipeline, Pipeline Shield, Derivo and P64 flow diverters in an in vitro pulsatile flow human blood loop model

Gaurav Girdhar, Samantha Ubl, Reza Jahanbekam, Sinduja Thinamany, A. M. Belu, J Wainwright, Michael Wolf

eNeurologicalSci · 2019 · 46 citations · 25 references

DOIFull text

Open access

Concepts

Abstract

). High resolution scanning electron microscopy performed on the intraluminal and cross-sectional surfaces of each device showed the lowest accumulation of platelets and fibrin on Pipeline Shield relative to Derivo, P64, and Pipeline. Derivo and P64 also had higher thrombus accumulation at the flared ends. Pipeline device with Phosphorylcholine surface treatment (Pipeline Shield) could mitigate device material related thromboembolic complications.

References

25