Application of <scp>TG</scp>‐100 risk analysis methods to the acceptance testing and commissioning process of a Halcyon linear accelerator

24

Citations

16

References

2019

Year

Abstract

This paper presents the results of FMEA and quality control measures to mitigate the FMs for the ATC process for Halcyon machine. Unique FMs that result from the differences in the ATC guidelines provided by the vendor and current conventional protocols, and the challenges of performing the ATC due to the new linac features and ring-gantry design were highlighted for the first time. The FMs identified in the present work along with the suggested quality control measures, could potentially be used to improve the design features of future ring-gantry type of linacs that are likely to be preassembled, preconfigured, and heavily reliant on automation and image guidance.