Evaluation of Photodiode-based Pyranometers and Reference Solar Cells on a Two-Axis Tracking System

Frank Vignola, Josh Peterson, Rich Kessler, Mike Dooraghi, Manajit Sengupta, F. Mavromatakis

2018 · 14 citations · 4 references

Concepts

Abstract

The performance of photodiode-based pyranometers and reference solar cells are examined on a two-axis tracker surface. This is the third in a series of papers examining the performance of photodiode-based pyranometers and reference solar cells on fixed and tracking surfaces. Because the instruments are oriented normally to the incident radiation, angle-of-incidence effects are minimized. By removing the angle-of-incidence effects inherent in earlier studies, the change in the sensor's clear sky responsivity can be more clearly identified. It was found that the photodiode-based pyranometers demonstrate a larger change in responsivity than the reference solar cells. Spectral measurements are recommended to verify the cause of this difference.

References

4