Nuclear Technology · 2018 · 11 citations · 18 references
EngineeringElectrical EngineeringOptoelectronicsPhotonicsRadiation DetectionDetector TechnologyPhotodetectorsRadiation Imaging DetectorsScintillation DetectorsScintillator MaterialsPhotomultiplier TubesPlastic ScintillatorsPosition SensitivityPosition-sensitive DetectorsReadoutReadout InstrumentationScintillation DetectionScintillator Detector Resolution
This work presents the design of a position-sensitive detector that we are evaluating for cosmic-ray muon imaging. The position-sensitive detector consists of an EJ-200 plastic scintillator panel that is 32 × 32 × 2.5 cm in dimension. A quantity of 32 parallel grooves, each 2 mm wide and 4 mm deep with a pitch of 1 cm, are carved on the top and bottom sides, in perpendicular orientation, of a scintillator panel. Two wavelength shifting optical fibers are embedded in each groove for light collection and transport. The optical fibers from each channel are coupled to one pixel of a Hamamatsu H8500C multi-anode photomultiplier tube. An encoding technique using a one-dimensional resistor network was developed to reduce the number of required readout channels for position determination. The position calibration was performed with a blue light emitting diode. The detector was shown to achieve position resolution of ~1 cm (sigma).
18
S. Eidelman, K. Hayes, Keith A. Olive et al. · Physics Letters B · 2004 · 6.7K citations
Natural Sciences, Physics, Cosmology +16
G.P. Yost, R. M. Barnett, I. Hinchliffe et al. · Physics Letters B · 1988 · 4.4K citations
Review of particle physics. Particle Data Group
S. Eidelman, G. Höhler, R. Sekhar Chivukula et al. · arXiv (Cornell University) · 2004 · 2.4K citations · Full text