Assessment of drug delivery devices

Elsa Batista, Nelson Almeida, Andreia Furtado, Eduarda Filipe, Luís de Sousa, Rui F. Martins, Peter Lucas, Harm Tido Petter, Roland A. Snijder, Annemoon Timmerman

Biomedizinische Technik/Biomedical Engineering · 2015 · 33 citations · 6 references

DOIFull text

Open access

Concepts

Abstract

For critical drug delivery, it is important to have a constant and well-known infusion rate delivered by the complete infusion set-up (pump, tubing, and accessories). Therefore, various drug delivery devices and accessories were tested in this article in terms of their infusion accuracy, start-up delay, response time, and dependency on the viscosity. These measurements were performed as part of the European funded research project MeDD. The obtained results show that the infusion accuracy of the devices is flow rate and accessory depended, especially for low flow rates. Viscosity does not have a significant impact on the flow rate accuracy.

References

6