Journal of Applied Clinical Medical Physics · 2018 · 18 citations · 20 references
MedicineRadiation OncologyRadiologyDiagnostic ImagingDiagnostic RadiologyTreatment PlanningComputed TomographyImage QualityImage Quality AssessmentMedical ImagingRadiographic ImagingStereotactic RadiosurgeryComputed Tomography PhantomsCone-beam Computed Tomography (Cbct)Image Acquisition ParametersLinear Accelerator RadiosurgeryPhantom ValidationRadiographic Image QualityStereotactic Radiosurgery DosimetryTomographic Image Quality
PURPOSE: We explore the optimal cone-beam CT (CBCT) acquisition parameters to improve CBCT image quality to enhance intracranial stereotactic radiosurgery (SRS) localization and also assess the imaging dose levels associated with each imaging protocol. METHODS: linear accelerator (Varian Medical Systems, Palo Alto, CA) with different x-ray tube current and potential settings, gantry rotation trajectories, and gantry rotation speeds. To assess image quality, images of the Catphan 504 phantom were analyzed to evaluate the following image quality metrics: uniformity, HU constancy, spatial resolution, low contrast detection, noise level, and contrast-to-noise ratio (CNR). To evaluate the imaging dose for each protocol, the cone-beam dose index (CBDI) was measured. To validate the phantom results, further analysis was performed with an anthropomorphic head phantom as well as image data acquired for a clinical SRS patient. RESULTS: The Catphan data indicates that adjusting acquisition parameters had direct effects on the image noise level, low contrast detection, and CNR, but had minimal effects on uniformity, HU constancy, and spatial resolution. The noise level was reduced from 34.5 ± 0.3 to 18.5 ± 0.2 HU with a four-fold reduction in gantry speed, and to 18.7 ± 0.2 HU with a four-fold increase in tube current. Overall, the noise level was found to be proportional to inverse square root of imaging dose, and imaging dose was proportional to the product of total tube current-time product and the cube of the x-ray potential. Analysis of the anthropomorphic head phantom data and clinical SRS imaging data also indicates that noise is reduced with imaging dose increase. CONCLUSIONS: Our results indicate that optimization of the imaging protocol, and thereby an increase in the imaging dose, is warranted for improved soft-tissue visualization for intracranial SRS.
20
Evaluation of on-board kV cone beam CT (CBCT)-based dose calculation
Yong Yang, Eduard Schreibmann, Tianfang Li et al. · Physics in Medicine and Biology · 2007 · 293 citations
A dose comparison study between XVI<sup>®</sup>and OBI<sup>®</sup>CBCT systems
William Y. Song, Srijit Kamath, Shuichi Ozawa et al. · Medical Physics · 2008 · 143 citations
Imaging doses from the Elekta Synergy X-ray cone beam CT system
Ali Amer, Tom Marchant, Jonathan Sykes et al. · British Journal of Radiology · 2007 · 138 citations · Full text