Publication | Open Access
Correlations Between Corneal Biomechanical Properties Measured With the Ocular Response Analyzer and ICare Rebound Tonometry
68
Citations
19
References
2008
Year
IOP values obtained with the rebound tonometer are higher in thicker corneas and are positively correlated with biomechanical corneal parameters, namely corneal resistance factor. Although corneal thickness plays a significant role in rebound tonometry, elastic and viscous properties of the cornea seem to play a significant role in the interaction of the tonometer probe with the ocular surface. However, the mechanism behind this process is presently unknown.