Journal of Circuits Systems and Computers · 2018 · 11 citations · 5 references
EngineeringBiomedical EngineeringElectrical EngineeringAnalog ElectronicsBiomedical Signal ProcessingLow-power DesignAnalog Integrated CircuitsCmos Circuit DesignDifferential AmplifiersLow-noise Circuit DesignNoise In Analog CircuitsNoise ReductionAnalog Circuit DesignAnalog Front-end CircuitsCmos Analog Integrated CircuitsInput-referred NoiseLow-noise AmplificationLow-noise AmplifiersOperational Transconductance AmplifiersSilicon
This paper presents a fully integrated front-end, low noise amplifier (LNA), dedicated to the processing of various types of bio-medical signals, such as Electrocardiogram (ECG), Electroencephalography (EEG), Axon Action Potential (AAP). A novel noise reduction technique, for an operational transconductance amplifier (OTA), has been proposed. This adds a current steering branch parallel to the differential pair, with a view to reducing the noise contribution by the cascode current sources. Hence, this reduces the overall input-referred noise of the LNA, without adding any additional power. The proposed technique implemented in 65[Formula: see text]nm CMOS technology achieves 30 dB closed-loop voltage gain, 0.05[Formula: see text]Hz lower cut-off frequency and 100 MHz 3-dB bandwidth. It operates at 1.2[Formula: see text]V power supply and draws 1[Formula: see text][Formula: see text]A static current. The prototype described in this paper occupies 3300[Formula: see text][Formula: see text]m 2 silicon area.
5
Shuang Song, M. J. Rooijakkers, Pieter Harpe et al. · IEEE Transactions on Biomedical Circuits and Systems · 2015 · 112 citations · Full text
Engineering, Biomedical Engineering, Electrical Engineering +17