Concepedia

Publication | Open Access

Development of large-volume high-performance monolithic CZT radiation detector

25

Citations

6

References

2018

Year

Abstract

Recent progress with Cadmium Zinc Telluride (CZT) radiation sensors grown by the traveling heater method (THM) at Kromek is reported. Large volume monolithic pixelated detectors, 40×40×15 mm<sup>3</sup> have been fabricated with good initial gamma spectroscopy response (< 2.5% energy resolution at 662 keV at room temperature without correction). After depth of interaction (DOI) correction, detector performance with < 1% energy resolution at 662 keV at room temperature has been obtained on pixelated 22×22×15 mm<sup>3</sup> CZT detector. For medical imaging applications, 20×20×6 mm<sup>3</sup> pixelated detectors exhibits < 3% energy resolution at 122 keV without correction. These results have been achieved via our proprietary THM crystal growth in combination with our robust device fabrication technique. Examples of progress in other areas of CZT development for gamma spectroscopy and imaging applications such as 40×40×5 mm<sup>3</sup> cross-strip device for PET and Kromek's general-purpose SPECT camera will also be presented.

References

YearCitations

Page 1