Clinical ophthalmology · 2018 · 14 citations · 23 references
Health ProfessionsOphthalmologySurgeryOcular ImagingOphthalmic SurgeryUltrasonographyAnterior Segment EvaluationAnterior Segment ImagingCataract AssessmentCataract EvaluationOcular BiometryOcular UltrasonographyIol Power CalculationLens BiometryPreoperative BiometrySignal-to-noise RatioSignal-to-noise Ratio AnalysisSurgical Planning
PURPOSE: To ascertain if optical biometry determination of axial length (AL) and intraocular lens (IOL) power is significantly different compared to ultrasound (US) biometry in cases with borderline signal-to-noise ratio (SNR). PATIENTS AND METHODS: Sixty patients who had cataract and IOL Master biometry with borderline SNR (1.6-2.0) were included. A retrospective chart review was performed to compare data collected with optical biometry and US biometry in cataract cases with borderline SNR. RESULTS: Results showed that optical biometry IOL and AL measurements were not significantly different from the US measurements. Analysis also demonstrated good agreement between the two methods. CONCLUSION: Our study suggests that, in cases of borderline quality data, IOL power and AL measurements with optical biometry are still useful in surgical planning and that additional US measurements may be used more as a corroborative tool.