Concepedia

Publication | Closed Access

Evaluation of patient and staff exposure with state of the art X‐ray technology in cardiac catheterization: A randomized controlled trial

10

Citations

29

References

2018

Year

Abstract

The new x-ray and image processing technology, significantly reduces patient dose in coronary angiographies, and PCIs by 67%. In general, scatter dose was also reduced, yet for some dosimeters the reduction was limited and not statistically significant. This study clearly indicates that the scatter behavior is highly dependent on C-arm rotation, operator movement and height, dosimeter position, beam filtration, clinical procedure type and system geometry.

References

YearCitations