Medical Physics · 2018 · 20 citations · 21 references
MedicineRadiologyBiomedical ImagingRadiation DosimetryRadiation MeasurementAbsorbed DosePatient-specific DosimetryPhantom DosimetryRadiation Dose MeasurementRadiation TherapyRadiographic ImagingBeam CollimationPatient Skin Dose MappingPatient Skin Dose MonitoringPhantom-based Dose MeasurementRadiation Dose AssessmentRadiation Dose EstimationRadiation Dose MappingX-ray Dosimetry
Purpose To perform a benchmark of a new DACS ‐integrated patient skin dose mapping solution using on‐phantom measurements with Gafchromic ® films. Materials and methods To calculate cumulative patient skin dose distribution with 1‐cm 2 resolution, a Radiation Dose Monitor ( RDM , Medsquare), using the Radiation Dose Structured Report ( RDSR ), tabulated backscatter and mass energy absorption coefficients together with site‐specific corrections for table, mattress attenuation, and air kerma calibration factor. Peak skin dose ( PSD ) and two‐dimensional (2D) skin dose distributions calculated with RDM were compared against on‐phantom measurements with XR ‐ RV 3 Gafchromic ® films considering two widely used x‐ray equipment. Seventeen different settings which include simple and multiple beam projections with extreme angulations (up to 75°), all available fields‐of‐view ( FOV s 48–11 cm), additional collimation, variable table height and lateral positions, and variable phantom thickness (12, 20, and 30 cm) were involved. Results Due to a careful calibration of films using clinical beam qualities, 22.8% (k = 2) overall measurement uncertainty was achieved. Calculated and measured PSD values agreed with an average difference of 10% ± 7% and 9% ± 7% for 34 test conditions performed on Siemens Artis Zee and GEMS Innova IGS interventional systems, respectively. Finally, RDM 's 2D skin dose maps closely matched those registered on XR ‐ RV 3 films considering the 1‐cm 2 resolution. While RDM correctly reproduced beam overlapping due to variable tube projections, FOV , table positions, etc., few challenges were identified related to conversion of rectangular fields to square areas in the RDSR and a stair‐step effect visible for large tube projections (>45°). Conclusion The accuracy of RDM 's DACS ‐integrated skin dose mapping software was acceptable considering measurement uncertainties associated with Gafchromic ® films.
21
International Commission on Radiological Protection
Radiology · 1969 · 2.2K citations
Skin dose mapping for fluoroscopically guided interventions
Perry Johnson, David Borrego, Stephen Balter et al. · Medical Physics · 2011 · 74 citations · Full text
et al. Hiroyuki Miyake · Japanese Journal of Radiological Technology · 2011 · 68 citations