Substituting gelatine with Pluronic F-127 matrix in 3D polymer gel dosimeters can improve nuclear magnetic resonance, thermal and optical properties

Malwina Jaszczak, Radosław A. Wach, Piotr Maras, Mariusz Dudek, Marek Kozicki

Physics in Medicine and Biology · 2018 · 49 citations · 55 references

DOIFull text

Open access

Concepts

Abstract

(VIC). The current research outcomes showed that not each polymer gel dosimeter could be manufactured with Pluronic F-127. Two of the polymer gel dosimeters (PAGAT and VIP) containing the Pluronic F-127 matrix allowed for some proper dose response for radiotherapy dosimetry (a response to a dose range of e.g. 0‒50 Gy). The new best performing Pluronic-based polymer gel dosimeters were characterised by improved nuclear magnetic resonance properties, when being compared to gels with gelatine matrix at the same monomer content. These are: (i) a ~33% higher dose sensitivity; (ii) a comparable or slightly higher linear and dynamic dose range and (iii) a lower (new VIP composition, VIP3) or equivocal (new PAGAT composition, PAGAT2-Pluronic) dose threshold. However, there might be optimised gelatine based polymer dosimeters demonstrating even better sensitivity. UV-vis spectrophotometry measurements revealed that Pluronic matrices ensure six-times lower (VIP3-Pluronic) and eight-times lower (PAGAT2-Pluronic) absorbance (at 400 nm) of non-irradiated gels compared to gelatine matrices, which makes the new polymer gel dosimeters optically improved in comparison to their corresponding gelatine-based compositions. The differences in absorption reduce for higher wavelengths. Differential scanning calorimetry measurements revealed the following temperature stability ranges for the gels: (i) VIP with gelatine matrix: 0 °C‒26 °C, (ii) VIP3 with Pluronic matrix: 13.8 °C-55.2 °C, (iii) PAGAT2 with gelatine matrix: 0 °C-80 °C and (iv) PAGAT2 with Pluronic matrix: 21.4 °C-55.2 °C. In conclusion, Pluronic F-127 is an attractive co-polymer to serve as a substitute for the gelatine matrix in some 3D polymer gel dosimeters.

References

55