MedicineOncologyRadiation OncologyRadiation DosimetryRadiation ProtectionTreatment PlanningExternal Beam RadiotherapyMonte Carlo DosimetryPhoton DosimetryRadiation TherapyRadiation Treatment PlanningReference DosimetryClinical Photon Beam DosimetryExternal Beam Radiotherapy DosimetryFlattening Filter Free Beam DosimetryFlattening Filter-free BeamsHigh-energy Photon Beam DosimetryHigh-energy Photon DosimetryParticle Beam DosimetryReference Field Dosimetry
BACKGROUND: The software PRIMO for the Monte Carlo simulation of radiotherapy linacs could potentially act as a independent calculation system to verify the calculations of treatment planning systems. We investigated the suitability of the PRIMO default beam parameters to produce accurate dosimetric results for 6 MV photon beams from Varian Clinac 2100 linacs and 6 MV flattening-filter-free photon beams from Varian TrueBeam linacs. METHODS: Simulation results with the DPM algorithm were benchmarked against a published reference dosimetry dataset based on point measurements of 25 dosimetric parameters on a large series of linacs. Studied parameters (for several field sizes and depths) were: PDD, off-axis ratios, and output factors for open fields and IMRT/SBRT-style fields. For the latter, the output factors were also determined with radiochromic film and with a small-sized ionization chamber. Benchmark data, PRIMO simulation results and our experimental results were compared. RESULTS: PDD, off-axis ratios, and open-field output factors obtained from the simulations with the PRIMO default beam parameters agreed with the benchmark data within 2.4% for Clinac 2100, and within 1.3% for TrueBeam. Higher differences were found for IMRT/SBRT-style output factors: up to 2.8% for Clinac 2100, and up to 3.3% for TrueBeam. Experimental output factors agreed with benchmark data within 1.0% (ionization chamber) and within 1.9% (radiochromic film). CONCLUSIONS: PRIMO default initial beam parameters for 6 MV photon beams from Varian Clinac 2100 linacs and 6 MV FFF photon beams from Varian TrueBeam linacs allowed agreement within 3.3% with a dosimetry database based on measurements of a high number of linacs. This finding represents a first step in the validation of PRIMO for the independent verification of radiotherapy plans.
17
Ioannis Sechopoulos, D. W. O. Rogers, Magdalena Bazalova‐Carter et al. · Medical Physics · 2017 · 287 citations · Full text