Concepedia

Publication | Open Access

A retrieval analysis of the Precice intramedullary limb lengthening system

37

Citations

22

References

2018

Year

Abstract

This study confirms less internal corrosion following modification, but evidence of titanium debris remains. We recommend no change to current clinical practice. However, potential reuse of the Precice nail, for secondary limb lengthening in the same patient, should be undertaken with caution.<b>Cite this article</b>: V. C. Panagiotopoulou, K. Davda, H. S. Hothi, J. Henckel, A. Cerquiglini, W. D. Goodier, J. Skinner, A. Hart, P. R. Calder. A retrieval analysis of the Precice intramedullary limb lengthening system. <i>Bone Joint Res</i> 2018;7:476-484. DOI: 10.1302/2046-3758.77.BJR-2017-0359.R1.

References

YearCitations