Publication | Open Access
A retrieval analysis of the Precice intramedullary limb lengthening system
37
Citations
22
References
2018
Year
This study confirms less internal corrosion following modification, but evidence of titanium debris remains. We recommend no change to current clinical practice. However, potential reuse of the Precice nail, for secondary limb lengthening in the same patient, should be undertaken with caution.<b>Cite this article</b>: V. C. Panagiotopoulou, K. Davda, H. S. Hothi, J. Henckel, A. Cerquiglini, W. D. Goodier, J. Skinner, A. Hart, P. R. Calder. A retrieval analysis of the Precice intramedullary limb lengthening system. <i>Bone Joint Res</i> 2018;7:476-484. DOI: 10.1302/2046-3758.77.BJR-2017-0359.R1.